The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster™

Kwok Cheung Andrew Lam, Rufina Chan, Peter C K Pang

Research output: Journal article publicationJournal articleAcademic researchpeer-review

138 Citations (Scopus)

Abstract

Background: Axial length and anterior chamber depth have been measured clinically using conventional ultrasound biometry. Recently, a non-contact device has become available to measure these parameters. This study evaluated the repeatability and accuracy of this device. Methods: The axial length and anterior chamber depth were measured by two practitioners on a group of young subjects using the IOLMaster™ followed by a conventional ultrasound biometer operated by a third practitioner. The accommodation was controlled in ultrasound biometry through a full correction on the non-measured eye and a distant fixation target. Results: There was good repeatability and accuracy of axial length assessment. The mean difference between the IOLMaster™ and ultrasound biometry was -0.099 mm, with 95% limits of agreement between 0.66 and -0.85 mm. The axial length was slightly shorter from the IOLMaster™ and the difference was not significant. The anterior chamber depth was repeatable but was shown to be deeper than the ultrasound results. The mean difference in anterior chamber depth between the IOLMaster™ and ultrasound biometry was 0.15 mm, with 95% limits of agreement between 0.34 and -0.03 mm. It is suggested that the former device is not measuring the axial anterior chamber depth. Conclusions: The IOLMaster™ is a non-contact 'optical' A-scan which is simple to use and good for axial length assessment. The anterior chamber depth assessment should be further evaluated.
Original languageEnglish
Pages (from-to)477-483
Number of pages7
JournalOphthalmic and Physiological Optics
Volume21
Issue number6
DOIs
Publication statusPublished - 10 Nov 2001

ASJC Scopus subject areas

  • Ophthalmology
  • Sensory Systems
  • General Health Professions

Fingerprint

Dive into the research topics of 'The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster™'. Together they form a unique fingerprint.

Cite this