T1 - Repeatability of corneal biomechanics waveform signal parameters derived from Ocular Response Analyzer in children
AU - Wan, Kin
AU - Cheung, Sin Wan
AU - Wolffsohn, James Stuart
AU - Cho, Pauline
N1 - Funding Information:
K. Wan is supported by the Research Residency Scheme of The Hong Kong Polytechnic University .
Publisher Copyright:
© 2020 British Contact Lens Association
Copyright:
Copyright 2020 Elsevier B.V., All rights reserved.
PY - 2020/10/23
Y1 - 2020/10/23
N2 - Purpose To investigate the repeatability of waveform signal parameters, measured with the Ocular Response Analyzer (ORA), in children. Methods Two sets of ORA measurements, with a 10-min break between them, were performed on children, aged six to 0.90 (regarded as moderate to excellent agreement). Results A total of 30 participants [15 SVS, 15 ortho-k (3.6 ± 2.4 months)] completed the study. Since no significant between-group differences were detected in demographic data (p > 0.28) and all waveform signal parameters (p > 0.05), data from the two groups of participants were pooled for the analysis of repeatability. Six parameters, h2, h21, p1area, p1area1, p2area, and p2area1, achieved ICCs (95 % CI) of 0.82−0.85 (0.61−0.93). The mean (SD) of these six parameters were 372 (91), 248 (61), 4077 (854), 1762 (399), 2359 (670), and 1020 (300), respectively. Bland-Altman plots and 95 % limits of agreement (95 % LoA) showed considerable agreement for all six parameters, the mean difference (95 % LoA) were -3 (-101 to 94), -2 (-67.56–62.70), 111 (-723 to 946), 102 (-334 to 539), 25 (-718 to 768), and -3 (-350 to 343), respectively. Conclusions Six waveform signal parameters (h2, h21, p1area, p1area1, p2area, and p2area1), which represent or are related to the areas under the waveform at the peaks in the signal, had moderate to excellent agreement in children. Results of the current study provides fundamental information for further studies on the potential clinical application of these waveform signal parameters in children.
AB - Purpose To investigate the repeatability of waveform signal parameters, measured with the Ocular Response Analyzer (ORA), in children. Methods Two sets of ORA measurements, with a 10-min break between them, were performed on children, aged six to 0.90 (regarded as moderate to excellent agreement). Results A total of 30 participants [15 SVS, 15 ortho-k (3.6 ± 2.4 months)] completed the study. Since no significant between-group differences were detected in demographic data (p > 0.28) and all waveform signal parameters (p > 0.05), data from the two groups of participants were pooled for the analysis of repeatability. Six parameters, h2, h21, p1area, p1area1, p2area, and p2area1, achieved ICCs (95 % CI) of 0.82−0.85 (0.61−0.93). The mean (SD) of these six parameters were 372 (91), 248 (61), 4077 (854), 1762 (399), 2359 (670), and 1020 (300), respectively. Bland-Altman plots and 95 % limits of agreement (95 % LoA) showed considerable agreement for all six parameters, the mean difference (95 % LoA) were -3 (-101 to 94), -2 (-67.56–62.70), 111 (-723 to 946), 102 (-334 to 539), 25 (-718 to 768), and -3 (-350 to 343), respectively. Conclusions Six waveform signal parameters (h2, h21, p1area, p1area1, p2area, and p2area1), which represent or are related to the areas under the waveform at the peaks in the signal, had moderate to excellent agreement in children. Results of the current study provides fundamental information for further studies on the potential clinical application of these waveform signal parameters in children.
KW - Corneal biomechanics
KW - Myopia
KW - Orthokeratology
UR - http://www.scopus.com/inward/record.url?scp=85093932198&partnerID=8YFLogxK
U2 - 10.1016/j.clae.2020.10.003
DO - 10.1016/j.clae.2020.10.003
M3 - Journal article
SN - 1367-0484
VL - 44
JO - Contact Lens and Anterior Eye
JF - Contact Lens and Anterior Eye
IS - 5
M1 - 101373
ER -