Abstract
Oxygen sensors prepared from tetramethyl orthosilicate and dimethoxy dimethylsilane with tris(4,7-diphenyl-1,10-phenanthroline)ruthenium (II) as the sensing dye were coated onto the well bottom surface of a 96-well polystyrene microtiter plate to give a high-throughput system for dissolved oxygen measurement. The oxygen sensors give linear Stern-Volmer calibration plots, and produce reliable and reproducible results in the determination of IC50values of drugs on a yeast model.
Original language | English |
---|---|
Pages (from-to) | 120-126 |
Number of pages | 7 |
Journal | Sensors and Actuators, B: Chemical |
Volume | 123 |
Issue number | 1 |
DOIs | |
Publication status | Published - 10 Apr 2007 |
Keywords
- High-throughput
- Microtiter plate
- Optical
- Oxygen
- Sol-gel
ASJC Scopus subject areas
- Electronic, Optical and Magnetic Materials
- Instrumentation
- Condensed Matter Physics
- Surfaces, Coatings and Films
- Metals and Alloys
- Electrical and Electronic Engineering
- Materials Chemistry