Evaluation of optimum room entry times for radiation therapists after high energy whole pelvic photon treatments

Lavine Ho, Peter White, Edward Chan, Kim Chan, Janet Ng, Timothy Tam

Research output: Journal article publicationJournal articleAcademic researchpeer-review

8 Citations (Scopus)

Abstract

Objectives: Linear accelerators operating at or above 10 MV produce neutrons by photonuclear reactions and induce activation in machine components, which are a source of potential exposure for radiation therapists. This study estimated gamma dose contributions to radiation therapists during high energy, whole pelvic, photon beam treatments and determined the optimum room entry times, in terms of safety of radiation therapists. Methods: Two types of technique (anteriorposterior opposing and 3-field technique) were studied. An Elekta Precise treatment system, operating up to 18 MV, was investigated. Measurements with an area monitoring device (a Mini 900R radiation monitor) were performed, to calculate gamma dose rates around the radiotherapy facility. Measurements inside the treatment room were performed when the linear accelerator was in use. The doses received by radiation therapists were estimated, and optimum room entry times were determined. Results: The highest gamma dose rates were approximately 7 μSv/h inside the treatment room, while the doses in the control room were close to background (~0 μSv/h) for all techniques. The highest personal dose received by radiation therapists was estimated at 5 mSv/yr. To optimize protection, radiation therapists should wait for up to11 min after beam-off prior to room entry. Conclusions: The potential risks to radiation therapists with standard safety procedures were well below internationally recommended values, but risks could be further decreased by delaying room entry times. Dependent on the technique used, optimum entry times ranged between 7 to 11 min. A balance between moderate treatment times versus reduction in measured equivalent doses should be considered.
Original languageEnglish
Pages (from-to)131-140
Number of pages10
JournalJournal of Occupational Health
Volume54
Issue number2
DOIs
Publication statusPublished - 7 Aug 2012

Keywords

  • High energy photons
  • Radiographers
  • Room entry times

ASJC Scopus subject areas

  • Public Health, Environmental and Occupational Health

Fingerprint

Dive into the research topics of 'Evaluation of optimum room entry times for radiation therapists after high energy whole pelvic photon treatments'. Together they form a unique fingerprint.