Comparison of the performance of the Nidek NT-2000 noncontact tonometer with the Keeler Pulsair 2000 and the Goldmann applanation tonometer

Hie Hua Wong, Thomas Lui

Research output: Journal article publicationJournal articleAcademic researchpeer-review

43 Citations (Scopus)

Abstract

We compared the performance of a new noncontact tonometer (NCT), Nidek NT-2000, with the Keeler Pulsair 2000 and the Goldmann applanation tonometer (AT). Twenty-two subjects with intraocular pressure (IOP) ranging from 10 to 21 mmHg were recruited and their IOP measured using each technique on 2 different days. The NT-2000 produces repeatable IOP readings and is comparable to the Goldmann AT. There is a significant between-visits difference in IOP when using the Pulsair 2000. The reproducibility of IOP measurements made with the Pulsair 2000 is relatively poor when compared to the Goldmann AT and the NT-2000. IOP measurements made with the Goldmann AT show good repeatability and reproducibility when the mean of three IOP readings are used. When a single Goldmann reading is used variability is increased. There is no significant difference in the IOP readings made on the first visit between techniques.
Original languageEnglish
Pages (from-to)51-58
Number of pages8
JournalOptometry and Vision Science
Volume74
Issue number1
DOIs
Publication statusPublished - 1 Jan 1997

Keywords

  • applanation tonometer
  • intraocular pressure
  • noncontact tonometer
  • repeatability
  • reproducibility

ASJC Scopus subject areas

  • Ophthalmology

Fingerprint

Dive into the research topics of 'Comparison of the performance of the Nidek NT-2000 noncontact tonometer with the Keeler Pulsair 2000 and the Goldmann applanation tonometer'. Together they form a unique fingerprint.

Cite this