Abstract
This paper presents formulations to evaluate noise in differential and ratiometric measurements that are often performed in biosensing. These measurements are performed to improve signal to noise ratio of the sensing systems for sensitive detection of dynamic biological processes. The use of these formulations is discussed in the context of the differential intensity surface plasmon resonance (SPR) system that is widely used to characterise molecular interactions on a confined axial scale. Previous studies provide qualitative descriptions of the noise performance of such systems but lack rigorous characterisation. Here we present analytical expressions for quantitative evaluation of the noise in differential and ratiometric measurements by applying the rules of arithmetic operations on random variables. Such formulations provide the means for evaluating the signal to noise ratio of such systems. We present how correlated noise can be removed by performing differential or ratiometric processing. Applying these formulations, we also show how the sensitivity of the differential intensity SPR system changes during the experiment.
Original language | English |
---|---|
Pages (from-to) | 1059-1067 |
Number of pages | 9 |
Journal | Sensors and Actuators, B: Chemical |
Volume | 260 |
DOIs | |
Publication status | Published - 1 May 2018 |
Keywords
- Biosensing
- Chemical sensors
- Differential or ratiometric measurements
- Surface plasmon resonance sensors
ASJC Scopus subject areas
- Electronic, Optical and Magnetic Materials
- Instrumentation
- Condensed Matter Physics
- Surfaces, Coatings and Films
- Metals and Alloys
- Electrical and Electronic Engineering
- Materials Chemistry